The Global Calibration Initiative uses measurements from the international AFM community to perform non-invasive calibration of AFM cantilevers (of any geometry) with the Sader Method. A live database lets users instantly compare their calibration measurements to those of others in the community, while systematically refining their own calibration using others' data. The GCI provides a reference point for worldwide calibration, enabling force measurements from different labs to be compared directly.
NOTE: Take measurements of the fundamental flexural mode in air - with your cantilever at least 100 µm from the surface.
The GCI reports the static normal spring constant at the imaging tip position.
The spring constant, \(k\), is evaluated using the Sader Method (Ref. 1):
\(k_\mathrm{Sader}= \displaystyle A \, Q\, f_R^{1.3}\)
which requires only the resonant frequency, \(f_R\), and quality factor, \(Q\), of the uncalibrated cantilever in air, measured far from any surface.
The coefficient, \(A\), is universal for a particular cantilever geometry, regardless of cantilever thickness or coatings, and is obtained from measurements of a reference cantilever:
\(A = \displaystyle \frac{k_\mathrm{ref}}{Q_\mathrm{ref} \, f_{R, \mathrm{ref}}^{1.3}}\)
where the subscript 'ref' refers to the reference cantilever. Multiple reference cantilevers are used to determine the \(A\)-coefficient (Ref. 2):
\(A= \displaystyle \frac{1}{N} \sum_{i=1}^{N} A_i = \frac{1}{N} \sum_{i=1}^{N} \frac{k_{\mathrm{ref}, i}}{Q_{\mathrm{ref}, i} \, f_{R, \mathrm{ref}, i}^{1.3}}\)
where \(N\) is the total number of measurements and the subscript \(i\) refers to an individual measurement. This average systematically reduces uncertainty in the \(A\)-coefficient which is used to standardise calibration.
The Global Calibration Initiative utilises measured data from the international AFM community to determine the \(A\)-coefficient - seamlessly enabling the standardisation and non-invasive calibration of any cantilever type (which users can specify*).
This is performed from users' measurements of \(f_{R, \mathrm{ref}}\), \(Q_\mathrm{ref}\) and \(k_\mathrm{ref}\) in air obtained using the Thermal Method (which is built into most AFMs).
A histogram of the individual \(A\)-coefficients from users is provided, along with the total number of data samples and users for each cantilever.
Details of the Global Calibration Initiative are in Ref. 3.
* Users can suggest a cantilever type for the Global Calibration Initiative by clicking on the "Request Cantilever" button.
References